Typical applications | In order to protect personnel from exposure to dangerous levels of radiated EMF it is not advisable to measure electromagnetic field strength in certain areas. Here is a short list of situations/locations where this instrument may be useful: • Any RF field strength measurement • Mobile phone base station antennae • Wireless communication applications (TDMA, GSM, CW, DECT) • RF Power measurement for transmitters • Detection and installation of wireless LAN (WiFi) • Detection and location of Spy Cameras and Wireless Bugs • Measuring radiated EMF from cellular/cordless phones • Microwave leakage detection • Detecting electromagnetic radiation in personal living environment |
---|---|
Frequency range | 50 MHz to 3.5 GHz |
Specified measurement range: CW signal (freq. > 50 MHz) |
20mV/M to 108.0V/m |
53μA/m to 286.4mA/m | |
1μW/m2 to 30.93W/m2 | |
0μW/cm2 to 3.093mW/cm2 | |
Dynamic range | 75 dB typical |
Isotrophy deviation | Typically ±1.0 dB for frequency > 50 MHz |
Frequency response | ±1.0 dB (50 MHZ to 1.9 GHz) ±2.4 dB (1.9 GHz to 3.5 GHz) |
Absolute error | ±1.0 dB (at 1V/m and 50 MHz) |
Overload limit | 10.61mW/cm2 (200 V/m) |
Thermal response (0 to 50°C) | ±0.2dB |
Measurement method | Digital, triaxial measurement |
---|---|
Directional characteristics | Isotrophic, triaxial |
Manual Data Memory/Read | 99 sets of data memory and read storage |
Measurement range selection | One continuous range |
Display values | Instantaneous measured value, maximum value, or maximum average value |
Display resolution | 0.1mV/m, 0.1μA/m, 0.001μW/m2, 0.001μW/cm2 |
Operating temperature | 0 to 50°C, (32 to 122°F) |
Operating humidity | 25% RH to 75% RH |
Dimensions | 237(L) X 60(W) X 60(T)mm |
Weight | 200g |
Accessories | Instruction manual, Battery, Carrying Case |